HXK-1 programmable computer intelligent muffle furnace

NegotiableUpdate on 10/09
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

HXK-1 programmable computer intelligent muffle furnace

Product Details

HXK-1 programmable computer intelligent muffle furnace



Product Description of Programmable Computer Intelligent Muffle Furnace

Appearance: The HXK-1 programmable computer intelligent muffle furnace adopts an integrated structure of upper furnace body and lower temperature control, with a novel and beautiful appearance. The furnace door is open and easy to operate. Due to the integration of the furnace body and control part, the cumbersome wiring connection between the two parts on site is avoided, making it easy to move and saving operating space.

Main technical parameters of programmable computer intelligent muffle furnace:

1. Temperature control range: (0-1000) ℃, resolution: 1 ℃, accuracy: level 0.5.

Controllable time: programmable time (0-9999) min, resolution 1 min

Fixed program (0-99) min59s, resolution 1s

Time measurement accuracy ± 30s/24h

2. Load power: not exceeding 5KW

3. Working power supply: AC220V ± 20V/50Hz

4. Furnace size: 300 × 200 × 120 (mm)

5. Test items: slow ash, fast ash, volatile matter, Roga adhesion index, and self setting program

6. Dimensions: 500 × 620 × 770 (mm) (excluding chimney protrusion height)

Programmable computer intelligent muffle furnace product comes with built-in program:

(Complies with the national standard GB/T212-2001)

1. Slow ash content

2. Rapid ash content (815 ℃. 40min)

3. Rapid ash content (850 ℃ for 40 minutes)

4. Volatile matter (preheated to 900 ℃)

5. Volatile matter (preheated to 910 ℃)

6. Volatile matter (preheated to 920 ℃)

7. Volatile matter (preheated to 930 ℃)

8. Roga Bonding index (according to national standard GB5447-1985)

9. Programmable 1-3 constant temperature points

Common fault causes and troubleshooting methods of programmable computer intelligent muffle furnace

Fault code

cause of failure

exclusion method

Err1

System failure

1. Not at zero, check IC4051

2. The proportional coefficient is not around 950. Check IC7650 and its surrounding circuits;

3. The ambient temperature shows below zero. Adjust the 101 potentiometer and check it

LM134 (or 334)

Err2

Controllable silicon out of control

1. Thyristor breakdown, replace thyristor;

2. No trigger signal, check the trigger circuit;

Err3

Temperature display exceeds the range

This phenomenon is not a fault

Err4

Broken couple and temperature exceeding limit

1. Disconnect the thermocouple and replace it;

2. If the thermocouple compensation wire has poor contact or is disconnected, reconnect the thermocouple.

Err5

Temperature not sampled

1. Check IC4051; 2. Check IC7135.