Instrument filter integrity tester IntegtestV4.0 detailed introduction:
Membrane integrity testerTechnical Specifications:
| Power requirements/power | 170-240V AC, 50/60Hz; 110W |
| The maximum operating pressure | 8000mbar |
| Zui low intake pressure | 100 mbar |
| Overall dimensions | 240 (width) x 380 (depth) x 280 (height 1) x 220 (height 2) |
| test scope | 500-6900mbar |
| test accuracy | Net volume test: ± 4%; Bubble point: ± 50mbar; Diffusion flow: ± 4%; Water immersion: ± 0.1ml/min |
| operating conditions | Environmental temperature:+5 ℃~+35 ℃; Relative humidity: 10-80% |
| Test duration | Net volume test: 5 minutes ± 2 minutes; Diffusion flow test: 10 min ± 2 min; |
| Simple bubble point test | 15 min±2min; Enhanced bubble point test: 20 minutes ± 2 minutes; |
| Water immersion test | Set testing time+5 minutes ± 2 minutes; |
| Print function | Wide line Chinese printing, outputting test parameters, test results, and automatic printing function |
| History recording function | 500 sets of test results (including test curves), which can be increased to 1000 sets of test results according to user needs |
| display screen | Size: 5.7 'TFT; monochrome |
| serial connection | Serial port: RS232; |
| Language Options | Chinese, English |
Membrane integrity tester
The main features of IntegtestV4.0 instrument filter integrity tester are:
1. 5.7 "backlit digital LCD display with clear display;
2. Pure Chinese operating environment, in line with the usage habits of Chinese people;
3. Real time display of test data and curves, monitoring the entire testing process;
4. Dual core computing with accompanying upper computer software. Includes all functions of similar foreign instruments;
5. It has a more comprehensive user grading management function, and users can change their passwords as needed;
6. Added automatic protection function for the tested filter to avoid damage to the filter caused by overvoltage testing;
7. Provides settings for automatic printing function, making user operation simpler and more convenient;
8. We have designed a "water immersion method" testing function specifically for hydrophobic filter cartridges, which solves the problem that hydrophobic filter cartridges can only be tested for integrity after being immersed in organic solvents;
9. The Integtest V4.0 filter integrity tester adopts more precise components and more sophisticated algorithms, enabling the instrument to detect the diffusion flow of flat membrane filters, which is of great significance for judging the integrity and characteristics of the membrane.