E-mail
info@gti-test.com
-
Phone
17640053769
-
Address
No. 56-39 Huanghe North Street, Huanggu District, Shenyang City, COFCO Plaza 7 # 2212
Jitai Precision Instrument (Shenyang) Co., Ltd
info@gti-test.com
17640053769
No. 56-39 Huanghe North Street, Huanggu District, Shenyang City, COFCO Plaza 7 # 2212
3990-02 high-efficiency leak detector generator for biosafety cabinetsIt is a thermal and high-capacity aerosol generator. After injecting aerosol solution and turning on the machine, the internal heating element quickly transforms the liquid aerosol into a vapor state. With the assistance of inert gases (such as N2, CO2, etc.), the vapor is rapidly cooled to generate dispersed aerosol smoke output. Heating time as low as less than 2 minutes, high concentration aerosol output, widely used in filter leakage testing for small air purification units to large clean room air conditioning purification systems.
After the high-efficiency filter is put into operation, if there are leakage points or gaps due to insufficient installation, the predetermined purification effect cannot be achieved, and a leak test must be conducted on the high-efficiency filter. Due to the low concentration of atmospheric dust upstream of the tested high-efficiency filter, an aerosol generator needs to be used to supplement the dust to meet the testing conditions. By using a gas volume gel photometer to test the concentration of suspended particles upstream of the high-efficiency filter and conducting a uniform scanning test downstream, the leakage rate of the tested high-efficiency filter can be determined to determine whether there is leakage.
Application field
Efficient filter validation
Nuclear Energy and Fuel
Pharmaceutical industry, electronics industry
Medical operating room, clean room
Purification air conditioning system
Food processing, scientific experiments
GTI Aerosol Generator 3990-02
Product features and advantages
2.4-inch true color display screen, real-time monitoring of internal gas path pressure and heating zone temperature
High concentration aerosol output (500-70000 cfm)
Rapid heating system, heating time less than 2 minutes
The oil storage space has increased by 60% compared to similar products, ensuring long-term dust generation
Mechanical level gauge displays oil level and allows for more intuitive observation of liquid level
Suitable for multiple types of aerosols (PAO-4, DOS/DESH...)
Suitable for different air conditioning system flow rates
Portable dust generation solution
3990-02 can emit high concentration aerosols ranging from 500 to 70000 cfm, with a heating time of less than 2 minutes. It is widely used for filter leakage testing in small air purification units to large clean room air conditioning purification systems. If conducting tests on FFUs, biosafety cabinets, clean workbenches, etc. in the cleanroom, it is recommended to use Laskin Nozle aerosol generators 3990-01 or 3990-03.
3990-02 high-efficiency leak detector generator for biosafety cabinetsTechnical Specifications
Specifications | |
Model number |
3990-02 |
Mode of occurrence |
Heating occurs |
flow range |
833 - 119,000 m3/h (500 - 70,000 cfm) |
concentration range |
100mg/ m3@11 ,900 m3/h (100μg/l@7,000cfm) 10mg/ m3@119 ,000 m3/h (100μg/l@70,000cfm) |
heating time |
< 2min |
particle size distribution |
满足 ANSI/ASME N509/510 |
Aerosol solution |
PAO-4、DOS/DESH、 Ondina EL、DOP/DEHP、 mineral oil |
Capacity |
1.8L |
Power supply |
220VAC 50HZ 8A |
Input pressure |
345 kpa (50 PSI) |
Overall dimensions |
430 x 140 x360mm |
Weight |
11KG |
Flexible dust generation solution
Capable of producing industry standard polydisperse suspended particulate matter, and can adjust the concentration output according to actual needs to meet various application requirements. The 3990-02 series generator can be used in conjunction with the aerosol photometer 3990 to test the leakage of high-efficiency filters, facilitating the identification and quantification of leakage points for repair reference.
Durable design
3990-02 adopts a stainless steel shell, which will not deform due to excessive pressure and is easy to clean. Large capacity chamber, only needs to be adjusted to the appropriate temperature and pressure to work.
Nozzle connector
Optional nozzle connector can be used to direct high concentration dispersed suspended particles to a testing port located further away, ensuring no leakage.
2.4-inch true color display screen
Real time monitoring and display of internal gas path pressure and heating zone temperature, clear and accurate.
Service&Training&Maintenance @ GTI
Product calibration, maintenance, and fault repair
After sales service provided by partners trained and certified by GTI service system