E-mail
info@gti-test.com
-
Phone
17640053769
-
Address
No. 56-39 Huanghe North Street, Huanggu District, Shenyang City, COFCO Plaza 7 # 2212
Jitai Precision Instrument (Shenyang) Co., Ltd
info@gti-test.com
17640053769
No. 56-39 Huanghe North Street, Huanggu District, Shenyang City, COFCO Plaza 7 # 2212
GTI aerosol leak detection equipment for purification engineering 3990-03It is an aerosol generator that uses the Laskin Nozle method, has a built-in compressor, does not require any other gas source, and can work when powered on. Durable, portable and reliable. After injecting aerosol (PAO, etc.) solution, turning on the power, and adjusting the pressure to 20psi (0.14Mpa), when the flow rate is 50-2000cfm (85-3400m3/h), by adjusting Laskin Nozle, polydisperse suspended particles with a concentration of 10-100 μ g/L can be generated.
3990-03 is widely used in filter unit systems with flow rates not exceeding 2000cfm (3400m3/h), such as biosafety cabinets, laminar flow hoods, negative pressure filtration units, filtration modules, or movable clean units.
After the high-efficiency filter is put into operation, if there are leakage points or gaps due to insufficient installation, the predetermined purification effect cannot be achieved, and a leak test must be conducted on the high-efficiency filter. Due to the low concentration of atmospheric dust upstream of the tested high-efficiency filter, an aerosol generator needs to be used to supplement the dust to meet the testing conditions. By using an aerosol photometer to test the concentration of suspended particles upstream of the high-efficiency filter and conducting a uniform scanning test downstream, the leakage rate of the tested high-efficiency filter can be determined to determine whether there is leakage.
GTI Aerosol Generator 3990-03
Application field
Efficient filter validation
Nuclear Energy and Fuel
Pharmaceutical industry, electronics industry
Medical operating room, clean room
Biosafety cabinets, clean workbenches
Food processing, scientific experiments
......
Product features and advantages
Built in compressor, no need for other air sources, can work with power on
The concentration of suspended particulate matter can be adjusted from 10 to 100 μ g/L to meet different testing requirements
The flow range is adjustable from 50-2000cfm to meet the needs of different applications
Multiple aerosol solutions can be selected (PAO, DOP, Ondina EL...)
Stainless steel shell, pressure resistant design, no deformation due to excessive pressure
Large capacity chamber, ensuring long-term dust generation
......
GTI aerosol leak detection equipment for purification engineering 3990-03Technical Specifications
Model number |
3990-03 |
flow range |
50~2000cfm(85~3400m3/h) |
concentration range |
100 μ g/L @ flow rate 200cfm 10 μ g/L @ flow rate 2000cfm |
Aerosol solution |
PAO、DOP、 Multi dispersed (cold) |
Gas source |
Built in compressor |
Method of occurrence |
2 or 6 Laskin Nozzel |
Overall dimensions |
38(L)X27(W)X26(H)cm |
Weight |
18.5kg |
Power supply |
AC100~240V 50/60Hz |
Configuration |
Host, nozzle connector, operation manual |
Optional selection |
Portable protective case |
Portable dust generation solution
The 3990-03 series Laskin Nozle aerosol generator is equipped with a built-in compressor and does not require any other gas source. It can work when powered on. Durable, portable and reliable. Multi dispersed suspended particles with a concentration of 10-100 μ g/L can be generated. It is a dust generating equipment used for testing biological full cabinets, laminar flow hoods, negative pressure filtration units, filtration modules, or movable clean units. If dust needs to be generated in air conditioning ventilation systems for all or local end high-efficiency filters, it is recommended to use aerosol generator 3990-02 with a larger flow range.
performance guarantee
The product comes with a calibrated pressure gauge that monitors the pressure adjustment value in real-time, requiring only 20psi (0.14Mpa), meeting electrical safety requirements. The 3990-03 series generator can be used in conjunction with the aerosol photometer 3990 to test the leakage rate of high-efficiency filters. Convenient for identifying and quantifying leakage points, for repair reference. The concentration range can be achieved within 10-100 μ g/L, but the flow rate varies depending on the generator model.
easy to operate
3990-03 adopts a stainless steel shell, which will not deform due to excessive pressure and is easy to clean. Equipped with a solution visual window, oil mist discharge outlet, and stainless steel nozzle connector, it can guide dispersed suspended particles to a testing port at a farther location without leakage.
Portable protective case
Due to the heavy weight of the generator, a portable protective case is sometimes required when moving or transporting it. Internal scientific design, reasonable placement of relevant components.
Service&Training&Maintenance @ GTI
Product calibration, maintenance, and fault repair
Partners trained and certified by GTI service system
After sales service conducted
Thoughtful service, complete accessories, and timely repairs
GTI Aerosol Generator 3990-03