EMGA-921 typeHydrogen analyzerIt is the flagship model and user-friendly instrument for hydrogen analysis.
This is a high-precision, high repeatability hydrogen element analyzer suitable for material research and quality control in fields such as steel, new materials, catalysts, etc. This is a brand new product that has been optimized to meet the requirements of users.
The detection of hydrogen element is achieved by measuring hydrogen gas using thermal conductivity detection method.
Principle and Gas Flow Path Diagram
Place the sample in a graphite crucible, keeping the crucible between the upper and lower electrodes of the pulse furnace. A high current passes through the crucible to generate high temperature.
The hydrogen in the sample is released as hydrogen gas (H2), and its concentration is measured by a thermal conductivity detector (TCD).

high performance
Wide hydrogen detection range: 0-0.02%
Optimize the design of TCD detector for hydrogen detection.
●precision
SD ≤ 0.04 μ g/g or RSD ≤ 0.5%, whichever is met (reference gas)
●standard method
EMGA-921 complies with the requirements of various metal (titanium, zirconium, tantalum, etc.) analysis standard methods in JIS Z2614, JIS H1619, JIS H1664, JIS H0696, ASTM E1447.
Analysis example of JSS sample containing low concentration of hydrogen.

Humanized software
Detection interface
Humanized software makes operation exceptionally easy
The extracted gas signal is displayed in real-time in digital form, and is also displayed as a curve along with the temperature level. The release curve graph will be automatically saved.
In the measurement window, the sample weight can be automatically registered. The results are saved in a data table for easy management.

HORIBA's original maintenance navigator
Notify users when consumables need to be replaced to maintain the best condition of the equipment.
The method and steps for replacing parts described in the maintenance navigator are simple and easy to understand, and operators can carry out maintenance work without experience and technical reserves.

EMGA-921 typeHydrogen analyzer
principle
Hydrogen: Thermal conductivity detection method (TCD) for detecting hydrogen H2
Detection range *:
Hydrogen: 0-0.02%
*Reducing the sample size can expand the detection range
Sample size: 1.0 ± 0.1g
Sensitivity (minimum reading): Hydrogen: 0.001 ppm
Accuracy * * (reproducibility)
hydrogen
σ n-1 ≤ 0.04 ppm or RSD ≤ 0.5%, whichever is satisfied (reference gas)
**Based on a sample size of 1.0 ± 0.1g and the following automatic functions
display
Test result: LCD or printed output
Alarm information: LCD or print output
Flowchart: LCD
Heating furnace power
Pulse electrode furnace inert gas protection heating and melting
Heating furnace output power can be set within 0-8.0kW
Sample feeding: Dual feeding device for sample/flux
Automatic functions: automatic cleaning, automatic crucible loading
Integral conditions:
Pre set time points
Reaching a comparative level of integration
Pre set time points or points that reach a comparative level, whichever is shorter
Sample ID: Registered with up to 20 single byte characters
calibration
Single point or multi-point calibration (reference gas or standard sample)
Calibrate using previous analysis data
Calibration curve correction function
function
Real time display of release curve
Interrupt analysis
Self diagnosis and alarm display
Release curve analysis (overlay comparison, differential analysis, etc.)
Data output (RC-232C or TCP/IP)
Dimensions: 750mm (H) × 785mm (D) × 653mm (W)
The position of the sample feeding port is 650mm high from the desktop.
Weight: 230kg
When transporting, it can be divided into two parts (each part weighing less than 140kg)
Computer: A PC running the Windows 10 operating system
power supply
Host: AC200/220/230/240V ± 10%
Frequency: 50/60 Hz ± 1Hz
power consumption
Host: 12kVA (maximum)
Automatic cleaning device: 1.5kVA (maximum)
Grounding resistance: below 10 Ω
Installation requirements
Operating temperature: 5-40 ° C
Performance guarantee temperature: 5-35 ° C
Humidity: Maximum 80% RH between 5-31 ℃, linearly decreases to 50% RH between 31-40 ℃
Vibration: bidirectional amplitude of 20 microns, frequency band acceleration less than 0.098m/S2
Gas requirements
Carrier gas: argon gas, purity 99.995% or above, pressure 0.35MPa
Stainless steel pipe connection (O.D.3mm) and within 3m of the instrument
Power gas: dry air or nitrogen, pressure 0.45MPa
Nylon tube connection (O.D.6mm) and within 5m of the instrument
Cooling device: heat dissipation unit (independent external)
Electronic balance (optional)
An electronic balance that can display a minimum of 1-0.0mg online
Optional Automatic Voltage Regulator (AVR)
Capacity: 15kVA
Weight: 130kg