EMGA-920 typeOxygen/Nitrogen AnalyzerIt is the flagship model and user-friendly instrument for oxygen and nitrogen analysis.
This is a high-precision and high repeatability oxygen and nitrogen element analyzer, suitable for material research and quality control in fields such as steel, new materials, catalysts, etc. This is a brand new product that has been optimized to meet the requirements of users.
The detection of oxygen is achieved by measuring carbon monoxide and carbon dioxide using a non dispersive infrared detector, while nitrogen is measured using a thermal conductivity detector.
Place the sample in a graphite crucible, keeping the crucible between the upper and lower electrodes of the pulse furnace. A high current passes through the crucible to generate high temperature.
The oxide in the sample reacts with the graphite crucible to generate carbon monoxide (CO), which reaches the detector with the carrier gas. The released carbon monoxide gas can be directly analyzed after dust filtration, or non dispersive red detection method (NDIR) can be used to detect CO based on oxygen concentration, or CO2 generated by CO oxidation through an oxidant (copper oxide) can be detected.
The nitrogen element is released in the form of nitrogen during the heating process of the graphite crucible. After dust filtration and removal of CO2 and moisture, it is detected by a thermal conductivity detector.

Wide detection range and high detection accuracy
*Detection range: Oxygen: 0-5%, Nitrogen: 0-3%
*Detection accuracy: Standard sample: σ n-1 ≤ 0.3ppm or CV ≤ 1.0%
Reference gas: σ n-1 ≤ 0.02ppm or CV ≤ 0.5%
The detection of oxygen adopts dual detectors (CO and CO2), and the TCD detector used for nitrogen detection is a long cell type and has been optimized in design. Therefore, regardless of oxygen or nitrogen, a very wide detection range can be obtained.
Compared to traditional methods, signal processing optimization reduces electrical noise by half.

EMGA meets the requirements of analytical standard methods for steel, titanium, ceramics, etc
for example
*ISO10720, ISO17053
*JIS G1228
*ASTM E1019, ASTM E1569, ASTM E1409…
Humanized software makes operation exceptionally easy.
The extracted gas signal is displayed in real-time in digital form, and is also displayed as a curve along with the temperature level. The release curve graph will be automatically saved.
In the measurement window, the sample weight can be automatically registered. The results are saved in a data table for easy management.

Notify users when consumables need to be replaced to maintain the best condition of the equipment.
The method and steps for replacing parts described in the maintenance navigator are simple and easy to understand, and operators can carry out maintenance work without experience and technical reserves.

EMGA-920 typeOxygen/Nitrogen Analyzer
principle
Oxygen: Non dispersive infrared detection method
Nitrogen: Thermal Conductivity Test (TCD)
Detection range *:
Oxygen: 0-5%, Nitrogen: 0-3%
*Reducing the sample size can expand the detection range to 0-100%
Sample size: 1,0 ± 0,1g
Sensitivity (minimum reading): Oxygen/Nitrogen: 0.0000001% (m/m)
Accuracy * * (repeatability)
Oxygen/Nitrogen
σ n-1 ≤ 0.02% (m/m) or CV ≤ 0.5%, whichever satisfies (reference gas)
σ n-1 ≤ 0.3% (m/m) or CV ≤ 1.0%, whichever satisfies (standard sample)
**Based on a sample size of 1.0 ± 0.1g and the following automatic functions
display
1) Test result: LCD or printed output
2) Alarm information: LCD or print output
3) Flowchart: LCD
Heating furnace power
Pulse electrode furnace inert gas protection heating and melting
Heating furnace output power can be set within 0-8.0kW
Sample feeding: D sample/flux dual feeding device
Automatic functions: automatic cleaning, automatic crucible loading
Integral conditions:
Pre set time points
Reaching a comparative level of integration
Pre set time points or points that reach a comparative level, whichever is shorter
Sample ID: Registered with up to 20 single byte characters
calibration
1) Single point or multi-point calibration (reference gas or standard sample)
2) Calibrate using previous analysis data
3) Calibration curve correction function
function
1) Real time display of release curve
2) Interrupt analysis
3) Self diagnosis and alarm display
4) Release curve analysis (overlay comparison, differential analysis, etc.)
5) Data output (RC-232C or TCP/IP)
Dimensions: 750mm (H) × 785mm (D) × 653mm (W)
The position of the sample feeding port is 650mm high from the desktop
Weight: 230kg
When transporting, it can be divided into two parts (each part weighing less than 140kg)
Computer: A PC running the Windows 10 operating system
power supply
Host: AC200/220/230/240V ± 10%
Frequency: 50/60 Hz ± 1Hz
power consumption
Host: 12kVA (maximum)
Automatic cleaning device: 1.5kVA (maximum)
Grounding resistance: below 10 Ω
Installation requirements
Operating temperature: 5-40 ° C
Performance guarantee temperature: 5-35 ° C
Humidity: Maximum 80% RH between 5-31 ℃, linearly decreases to 50% RH between 31-40 ℃
Vibration: bidirectional amplitude of 20 microns, frequency band acceleration less than 0.098m/S2
Gas requirements
Carrier gas: helium, purity 99.995% or above, pressure 0.35MPa
Stainless steel pipe connection (O.D.3mm) and within 3m of the instrument
Power gas: dry air or nitrogen, pressure 0.45MPa
Nylon tube connection (O.D.6mm) and within 5m of the instrument
Cooling device: heat dissipation unit (independent external)
Electronic balance (optional)
Enable connection with electronic balance with a minimum display of 1-0.01mg. An electronic balance with a minimum display of 1-0.01mg can be connected online
Optional Automatic Voltage Regulator (AVR)
Capacity: 15kVA
Weight: 130kg