OX-TRAN 2/28 H, due to its unique 4-test chamber, can simultaneously test 4 barrier samples, with a throughput twice that of traditional OTR instruments. The new dual membrane horizontal testing chamber can be disassembled for sample preparation, and the user-friendly interface provides real-time comparison of testing status and target limits.
highest precision
Coulox ® The sensor is an absolute coulomb meter sensor on the market that does not require calibration
The detection limit is as low as 0.05 cc/(m2* day)
Fully automatic temperature and flow control simplifies testing and provides repeatability
High throughput
The industry's fastest sensor response time reduces overall testing time
The double membrane detachable test chamber allows for testing 4 samples at once, requiring less space
Automatic testing method, one click start testing
TruSeal nitrogen circulation seal can eliminate the need for absolute zero point in many tests, thereby reducing overall testing time
The OX-TRAN 2/28 H high-throughput analyzer can test the oxygen permeability (OTR) of high and low barrier membranes without the need for humidification. It uses MOCON's Coulox absolute value Coulombometer sensor, which does not require calibration and complies with ASTM D3985 standard. The analyzer ranges from 0.05 to 200 cc/(m2*The wide detection range of day provides very accurate and reproducible test results.
The 4 test chambers allow for the use or non use of reference films for testing in triplicate within the same test. Choosing to run known standard samples as controls can improve the credibility of the results. OX-TRAN 2/28 H has 4 test chambers, which are easy to operate and very suitable for quality control laboratories that perform a large number of OTR tests.
This analyzer has industry features such as a detachable horizontal testing chamber, fully automated testing, and can provide higher throughput in the laboratory. It also has a QC testing mode that can display real-time comparisons between test status and target limits. It can be equipped with a touch screen operation interface, and versions without a touch screen operation interface can be controlled through the local computer's color graphical interface.
For over 25 years, the MOCON OX-TRAN series instruments have been reliable and durable in the market. OX-TRAN 2/28 H is an excellent choice for conducting routine short penetration tests in busy laboratories.
Characteristics and advantages
Double membrane testing chamber box, doubling testing capability
Easily prepare samples on the workbench
One click release and loading
Pneumatic clamping ensures more reproducible results
Additional test chambers can be loaded and tested in stages
Eliminate individual zeroing
Shorten testing time (improve throughput)
More reproducible results
Store and call commonly used testing methods
One click start of complete testing cycle
Perform flow and temperature control throughout the entire testing cycle
Preprocessing mode to improve throughput
Printing and exporting test reports, including pass, fail, or warning test results
Sensor protection
The full-color graphical interface displays the comparison between real-time detection status and target limits
The instrument can be equipped with a color touch screen or connected to a PC
Intuitive automation operations require minimal training
Built in multi-level user management and data security
Event logs and troubleshooting guidelines
Shorten testing time, reduce overall testing time, and improve throughput
No need for sensor calibration
The lifespan of the sensor is longer than other sensors on the market
Compared to sensors that require calibration, the results are more accurate and reproducible
The sensor complies with ASTM D3985
12-month 100% parts and process warranty
Coulox Sensor 4-Year Warranty
unit |
OX-TRAN 2/28H |
OX-TRAN 2/28HR |
|
Test scope (50cm2) |
cc/(m2*Oh my god) |
0.05 to 200 |
0.05 to 200 |
cc/(100 in2*(Heaven) |
0.003 to 12.9 |
0.003 to 12.9 |
|
Test scope (5)cm2) |
cc/(m2*(Heaven) |
0.5 to 2000 |
0.5 to 2000 |
cc/(100 in2*Oh my god) |
0.03 to 129.0 |
0.03 to 129.0 |
|
Resolve (50cm)2) |
cc/(m2*(Heaven) |
0.0001 |
0.0001 |
repetitiveness |
cc/(m2*(Heaven) |
0.02 or 1%* |
0.02 or 1%* |
Test temperature range |
°C |
20 ° to 60 ° C ± 0.2 ° C |
20 ° to 40 ° C ± 0.2 ° C |
Controlled relative humidity testing range |
%Relative humidity |
0 |
0%, 50-90%+/-5%, only testing gases |
Film Size |
Not applicable |
4” x 4” std., 2” x 2” for 5.0 cm2 |
4” x 4” std., 2” x 2” for 5.0 cm2 |
film thickness |
Not applicable |
Up to 20 Mil for 50 cm2. Up to 120 Mil for 5.0+10 cm2 |
Up to 20 Mil is 50 cm2, up to 120 Mil is 5.0+10 cm2 |
Thin film testing area |
Not applicable |
50 cm2 standard, 10 cm2+5.0 cm2 elective |
50 cm2 standard, 10 cm2+5.0 cm2 elective |
*Whichever is larger
Physical specifications
15.5 inches high x 12 inches wide x 23 inches deep (39.4 cm x 30.4 cm x 58.0 cm)
95 pounds (43.1 kg)
Electrical requirements
Voltage: 100-240 VAC 50/60 Hz
Maximum power consumption: 700 VA
temperature |
operation |
22°C ± 2°C |
storage |
10 ° C to 30 ° C |
|
humidity |
operation |
20% to 80% RH (non condensing) |
storage |
5% to 85% RH (non condensing) |
carrier gas |
gas composition |
Nitrogen/hydrogen mixture (98%/2%) |
Rated gas supply pressure |
25 PSI, (1.72 bar), (172 kPa) |
|
Test gas |
gas composition |
100% oxygen (purity 99.9%) |
Rated gas supply pressure |
25 PSI, (1.72 bar), (172 kPa) |
Remote access and control (requires optional Perm Net Light software)
Standard Ethernet port (100T)
Dual USB 2.0 ports (for connecting keyboards, mice, or storage devices)
This product is certified by UL and CSA, uses the ETL mark of Intertek Group plc, and has passed CE certification.
Meets the following safety standards.
IEC 61010-1:2010 Third Edition
UL 61010-1:2012 Third Edition
CAN/CSA-C22.2 No. 61010-1:2012 Third Edition
Meets the following EMC standards.
EN 61326-1:2013
EN 61000-3-2:2006、A1:2009、A2:2009
EN 61000-3-3:2013
Compliant with the EU RoHS Directive 2011/65/EU.
accessories
Additional double membrane testing cabin
Certification film 1-3
face shield
PermWare Data Collection, Data Management, and Instrument Status Dashboard
21 CFR Part 11 User and Data Security
PC Remote Viewing Software - PermNet Lite
Basic Preventive Maintenance Plan
Advanced Preventive Maintenance Plan
application
packaging film
single layer
multi-layer
Coating film
ASTM D3985
ASTM F2622
ISO CD 15105-2
JIS K-7126-B
DIN 53380-3
GB/T 19789-2005
According toASTM D3985
Install the test sample as a semi barrier material between two chambers under ambient atmospheric pressure. One chamber is slowly purged with nitrogen gas flow, while the other chamber contains oxygen. When oxygen permeates through the membrane into the nitrogen carrier gas, it will be transported to the coulomb detector, where a current is generated that is proportional to the amount of oxygen entering the detector per unit time.