OX-TRAN 2/40 series packagingOxygen permeation analyzer
The OX-TRAN 2/40 series is the next generation improved version for conducting oxygen permeability (OTR) testing on overall packaging. These advanced analyzers integrate MOCON's mature OTR analyzer system and precise environmental chambers, taking the accuracy, repeatability, and ease of use of packaging testing to a whole new level. With special features designed specifically for packaging testing, they greatly shorten the overall testing time, while the design of the new packaging testing chamber makes it easy to prepare samples and improves the reproducibility of test results. They have 4 independent testing chambers with replaceable test chamber boxes, which can conveniently and accurately test various sizes and types of packaging, many of which do not require the use of epoxy resin.
High precision packaging testing
●Coulox ® The sensor is an absolute coulomb meter sensor on the market that does not require calibration
The detection limit is as low as 25 * 10-7 cc/(pkg * day) (10X type)
Fully automatic temperature, flow, and relative humidity control simplifies testing and provides repeatability
easy to use
● Detachable test chamber, just touch one button to quickly exchange samples
● Different test chamber options make installation and testing of various packaging types simpler, more accurate, and reproducible
Advanced automatic packaging blowing and pre-processing functions can reduce overall testing time
The touch screen interface can control and monitor all test variables throughout the entire testing process
The Ox Tran 2/40 series integral packaging testing analyzer can test the oxygen permeation rate (OTR) of films and integral packaging from low to high barrier. They use Coulox absolute value Coulombometer sensors using MOCON, which do not require calibration and comply with ASTM D3985 standards. Four independent test chambers can conduct different tests simultaneously under precisely controlled environmental conditions (from 10 to 50 degrees Celsius, dry to 90% humidity).
We offer a variety of testing boxes that can conveniently and accurately test trays, bottles, flexible containers, small bags, cups, bottle caps, paper boxes, films, and other types of samples. Customized test chambers can also be used to simplify sample preparation and improve test repeatability.
Model:
This high-precision instrument series has two models, each with a different lower limit detection coefficient.
H-type (common barrier material) - commonly used detection range is 0.00025 to 1 cc/package * day
L-type (high barrier material) - capable of achieving a detection limit 10 times lower than H-type, with a detection range of 0.000025 to 1 cc/package * day
For over 25 years, the MOCON OX-TRAN series instruments have been reliable and durable in the market. The OX-TRAN 2/40 series is the latest choice that takes the entire packaging testing to a new level, enabling brand owners and packaging service providers to reliably test the shelf life of their finished packaging. It is a supplement to any testing laboratory that frequently tests complete packaging
Characteristics and advantages
Various packaging and film testing chambers
● Easily prepare samples on the workbench, many without the need for epoxy resin
One click quick release and loading
One time installation of foil simplifies the installation of packaging
A dedicated test chamber box can save time and improve the repeatability of testing
Automatic control function
● Flow, temperature, and relative humidity control
● Testing process development and recommendations
● Programmable testing sequence
Printing and exporting test reports
● Sensor protection
Touch screen interface
Intuitive automation operations require minimal training
Built in multi-level user management and data security
● Event logs and troubleshooting guidelines
No need to use a computer, saving valuable laboratory space
Coulox ® sensor
● Shorten testing time, reduce overall testing time, and improve throughput
No need for sensor calibration
The lifespan of the sensor is longer than other sensors on the market
Compared to sensors that require calibration, the results are more accurate and reproducible
The sensor complies with ASTM D3985
Advanced packaging testing function
● Automatically and quickly blow to the desired settings
● Automatic high flow regulation mode
Before testing begins, quickly check if the packaging is installed correctly
TruSeal®
● Eliminate individual zeroing
● Shorten testing time (improve throughput)
● More reproducible results
Install the test chamber without the need for tools
warranty
● 12-month 100% parts and process warranty
Coulox Sensor 4-Year Warranty
unit |
L-shaped |
H-shaped |
|
test scope |
cc/(m2x day) |
0.005 to 200 |
0.05 to 200 |
cc/(100 in2x day) |
0.0003 to 12.9 |
0.003 to 12.9 |
|
cc/(pkg x day) |
0.000025 to 1.0 |
0.00025 to 1.0 |
|
Compensate to 100% O2Scope of testing |
cc/(m2x day) |
0.025 to 1000 |
0.25 to 1000 |
cc/(100 in2x day) |
0.0015 to 64.5 |
0.015 to 64.5 |
|
cc/(pkg x day) |
0.000125 to 5.0 |
0.00125 to 5.0 |
|
resolution |
cc/(m2x day) |
0.002 |
0.02 |
reproducibility |
cc/(m2x day) |
0.002 or 1%* |
0.02 or 1%* |
*Whichever is larger
Ox Tran 2/40 series |
unit |
Test temperature range |
10 ° C to 50 ° C ± 0.2 ° C |
Controlled relative humidity testing range |
5 to 90% ± 4% |
Controlled relative humidity carrier gas range |
0 to 90% ± 4% |
Chamber size |
25.5 inches wide x 17 inches deep x 12 inches high |
Physical specifications
35.5 inches high x 33.0 inches wide x 29.75 inches deep (88.9 cm x 83.8 cm x 75.6 cm)
225 pounds (102 kg)
Electrical requirements
Voltage: 100-240 VAC 50/60 Hz
Maximum power consumption: 500 VA
temperature |
operation |
22°C ± 2°C |
storage |
10 ° C to 30 ° C |
|
humidity |
operation |
10% to 60% RH (non condensing) |
storage |
5% to 85% RH (non condensing) |
carrier gas |
gas composition |
Nitrogen/hydrogen mixture (98%/2%) |
Rated gas supply pressure |
25 PSI, (1.72 bar), (172 kPa) |
|
Test gas |
gas composition |
Dry compressed air (ISO 8753-1:2010 [1:3:1]) |
Rated gas supply pressure |
85 PSI, (5.86 bar), (586 kPa) |
Remote access and control (requires optional Perm Net Light software)
Standard Ethernet port (100T)
Dual USB 2.0 ports (for connecting keyboards, mice, or storage devices)
This product is certified by UL and CSA, uses the ETL mark of Intertek Group plc, and has passed CE certification.
Meets the following safety standards.
IEC 61010-1:2010 Third Edition
UL 61010-1:2012 Third Edition
CAN/CSA-C22.2 No. 61010-1:2012 Third Edition
Meets the following EMC standards.
EN 61326-1:2013
EN 61000-3-2:2006、A1:2009、A2:2009
EN 61000-3-3:2013
Compliant with the EU RoHS Directive 2011/65/EU.
accessories
80 cm2 packaging test chamber
Large vertical tray testing cabin
Soft packaging bag testing cabin
50 cm2 thin film testing chamber
Packaging adapter test chamber
Indoor regulating frame
Customized testing cabin
Certification film 1-4&set disposable packaging installation foil plate
21 CFR Part 11 User and Data Security
PC Remote Viewing Software - PermNet Lite
IQ/OQ files and verification
Basic Preventive Maintenance Plan
Advanced Preventive Maintenance Plan
application
packaging film
single layer
multi-layer
Coating film
Coated paper/cardboard
packaging
Ordinary food packaging
Medical packaging
Bottle/Cap
Test Standard
ASTM D3985
ASTM F2622
ASTM F1307
ASTM F1927
ISO CD 15105-2
JIS K-7126-B
DIN 53380-3
GB/T 19789-2005
GB/T 31354-2014
According to ASTM D3985
Install the test sample as a semi barrier material between two chambers under ambient atmospheric pressure. One chamber is slowly purged with nitrogen gas flow, while the other chamber contains oxygen. When oxygen permeates through the membrane into the nitrogen carrier gas, it will be transported to the coulomb detector, where a current is generated that is proportional to the amount of oxygen entering the detector per unit time.
OX-TRAN 2/40 series packagingOxygen permeation analyzer