X-ray fluorescence analyzer for sulfur in oil

NegotiableUpdate on 05/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The SLFA-6000 X-ray fluorescence oil sulfur analyzer has an optimized user interface, including options such as full-color display, USB output, data printing output mode, and a compact and refined appearance design. We have adopted a shutter shield to prevent X-ray leakage, as well as a new key switch function and a newly designed sample rotating table.
Product Details

SLFA-6000X-ray fluorescence analyzer for sulfur in oil

HORIBA Scientific announced the launch of a new SLFA-6100/6800 energy dispersive X-ray fluorescence sulfur analyzer, which measures sulfur content in oil ranging from 5 ppm to 9.9999%. This new, compact, and accurate analyzer meets any desired requirements.

The SLFA-6000 series features an optimized user interface, including options such as full-color display, USB output, data print output mode, and a compact and refined exterior design. We have adopted a shutter shield to prevent X-ray leakage, as well as a new key switch function and a newly designed sample rotating table.


Speed:

The measurement time can be set between 10 and 600 seconds.


Automation:

The SLFA-6800 is equipped with an 8-bit rotating sample stage, which can be used for analysis and testing under preset conditions.


Sampling operation:

After selecting the calibration curve and measurement conditions (or using "A" mode to select the correct calibration curve), simply press the "MEAS" button to start the measurement.


Accuracy:

This software allows saving up to 5 sets of 3 calibration curves to improve accuracy and usability.


SLFA-6000X-ray fluorescence analyzer for sulfur in oilconfiguration

Testing technical specifications

Detection Principle

X-ray fluorescence analysis method

Testing samples

Sulfur in petroleum products, such as heavy oil, naphtha, crude oil, and light oil

Testing Scope

5 ppm - 9.9999 %

repetitiveness

5ppm and below (samples containing 1% sulfur), 1.6ppm and below (samples containing 0% sulfur)

Error correction

Unit C/H within ± 50 ppm (for samples containing 1% sulfur)

detection limit

*3 times the standard deviation with a 0% sample of 5ppm or less (for 0 samples, 3 times the standard deviation)

Correction curve number

Up to 15 calibration curves

Correction curve order

Linear or 2-time (automatic or manual selection)

Sample quantity and quantity

4 - 10 mL, 1 sample (SLFA-6100)

4 - 10 mL, 8 samples (SLFA-6800)

Detection Time

10-600 seconds

data output

Print output, USB storage, USB interface connected to PC

spectral detection

Energy axis: 0-10keV, validated instrument performance through spectral analysis.

operating temperature

+ 5 ˚C to + 40 ˚C (+ 41 ˚F to 104 ˚F)

operating humidity

The maximum relative humidity within the range of 5 ° C to+30 ° C is 80%, and the maximum relative humidity decreases linearly to 50% within the range of 31 ° C to+40 ° C

power supply

AC 100 - 240 V +/- 10%, 50/60 Hz

power consumption

150 VA

Dimensions [W x D x H]

403 x 468 x 210 mm

weight

About 21 kg (SLFA-6100)

About 23 kg (SLFA-6800)


meet the standard

ASTM D4294 (USA), ISO 8754, JIS K2541/B7995